Accurate inference from quartz crystal microbalance (QCM) measurements in liquids is often limited by reducing resonance behavior to two scalar endpoints (frequency and dissipation shifts, and ) or by relying on single-equation analytical models (Kanazawa-model). We propose an AI-ready, impedance-resolved workflow that preserves full resonance line-shape information and converts it into compact, physically interpretable features for supervised regression. A passive microfluidic mixer generates glycerol–water concentration gradients under a constant total flow rate (50~L/min), while complex impedance spectra of a 10~MHz AT-cut quartz crystal are recorded in real time. Each sweep of nine spectra is parameterized via constrained Gaussian/Lorentzian models to yield 52 line-shape descriptors spanning extrema of , , and and peaks of , phase, and . The pipeline integrates consensus outlier handling, redundancy-aware feature ranking (mRMR), and cross-validated regression across linear, kernel, and ensemble models. Compared with the classical Kanazawa baseline, the impedance line-shape approach reduces concentration-prediction error from 0.456 to 0.148~%v/v RMSE (3.09 lower). The results demonstrate that impedance-resolved line-shape features provide a robust and interpretable basis for machine-learning-assisted QCM inference and illustrate a generalizable pattern for AI-enabled spectral sensing.