Coaxial 2-terminal bridge comparison is a widely used metrological method in the dissemination of value of capacitance up to 1000 pF and rad/s. In this range, capacitance can be measured with uncertainties as small as few parts in ; however, such measuments must take into account the corrections due to the impedance of cables, connectors and ports, which can tipically amount up to F/F and maybe larger. In this work, we report on how correction factors in 2-terminal pair coaxial bridge capacitance measurements are presently measured and calculated at Inmetro.