Photoplethysmography (PPG) signal comprises physiological information related
to cardiorespiratory health. However, while recording, these PPG signals are
easily corrupted by motion artifacts and body movements, leading to noise
enriched, poor quality signals. Therefore ensuring high-quality signals is
necessary to extract cardiorespiratory information accurately. Although there
exists several rule-based and Machine-Learning (ML) - based approaches for PPG
signal quality estimation, those algorithms’ efficacy is questionable. Thus,
this work proposes a lightweight CNN architecture for signal quality assessment
employing a novel Quantum pattern recognition (QPR) technique. The proposed
algorithm is validated on manually annotated data obtained from the University
of Queensland database. A total of 28366, 5s signal segments are preprocessed
and transformed into image files of 20 x 500 pixels. The image files are
treated as an input to the 2D CNN architecture. The developed model classifies
the PPG signal as good' or bad’ with an accuracy of 98.3% with 99.3%
sensitivity, 94.5% specificity and 98.9% F1-score. Finally, the performance of
the proposed framework is validated against the noisy `Welltory app’ collected
PPG database. Even in a noisy environment, the proposed architecture proved its
competence. Experimental analysis concludes that a slim architecture along with
a novel Spatio-temporal pattern recognition technique improve the system’s
performance. Hence, the proposed approach can be useful to classify good and
bad PPG signals for a resource-constrained wearable implementation.