Radioactive noble gas radon () is one of the major background sources below the MeV region in rare event search experiments. To precisely measure radon concentration in purified gases, a radon detector with an electrostatic collection method is widely used. In this paper, we discussed the improvements of a radon detector by installing a new PIN-photodiode ( ) whose surface area is times larger than that used previously ( ). We evaluated the detector’s performance by serially connecting two radon detectors equipped with two types of PIN-photodiodes. As a result of the calibrations, we found an improvement of in the detection efficiencies below , while a - improvement occurred above this level. The intrinsic background of the detector equipped with the large PIN-photodiode was measured as . This background level is consistent with the radon detector with the small PIN-photodiode, although we installed the large one. This improvement is useful for applications in radon emanation measurements from a material, which also emits water from its surface.