The performance of the energy degrader in terms of beam properties directly impacts the design and cost of cyclotron-based proton therapy centers. The aim of this study is to evaluate the performances of different existing and novel degrader materials. The quantitative estimate is based on detailed GEANT4 simulations that analyze the beam-matter interaction and provide a determination of the beam emittance increase and transmission. Comparisons between existing (aluminum, graphite, beryllium) and novel (boron carbide and diamond) degrader materials are provided and evaluated against semi-analytical models of multiple Coulomb scattering. The results showing a potential in emittance reduction for novel materials are presented and discussed in detail.

生成讲解博客的中文版本

生成AI智能校验的中文版本