加载中…
A highly granular electromagnetic calorimeter with scintillator strip readout is being developed for future lepton collider experiments. A prototype of 21.5 depth and mm transverse dimensions was constructed, consisting of 2160 individually read out mm scintillator strips. This prototype was tested using electrons of 2–32 GeV at the Fermilab Test Beam Facility in 2009. Deviations from linear energy response were less than 1.1%, and the intrinsic energy resolution was determined to be $(12.5 \pm 0.1 () \pm0.4 ())%/\sqrt{E[]}\oplus (1.2 \pm 0.1()^{+0.6}_{-0.7}())%$, where the uncertainties correspond to statistical and systematic sources, respectively.