We report an absolute energy response function to electronic and nuclear recoils for germanium and liquid xenon detectors. As a result, we show that the detection energy threshold of nuclear recoils for a dual-phase xenon detector can be a few keV for a given number of detectable photoelectrons. We evaluate the average energy expended per electron-hole pair to be 3.32 eV, which sets a detection energy threshold of 2.15 keV for a germanium detector at 50 mini-Kelvin at 69 volts with a primary phonon frequency of 1 THz. The Fano factors of nuclear and electronic recoils that constrain the capability for discriminating nuclear recoils below 2-3 keV recoil energy for both technologies are different.