Purpose: To demonstrate how diffusion blurs the longitudinal magnetization in steady-state gradient-echo imaging, quantify resolution dependence on flip angle and repetition time, and the effects on relaxometry at microscopic resolution. Methods: Diffusion was incorporated into the Bloch-Torrey equation in space to determine how the point spread function (PSF) of -weighted spoiled gradient-echo images varies with flip angle. Theoretical predictions were validated numerically and extended to b-SSFP. Results: Diffusion acts as a low-pass filter whose width depends on the diffusion rate (), repetition time (TR), and flip angle (). The blurring length spans from at large to as , equaling at the Ernst angle. Consequently, multi-angle acquisitions have mismatched resolution (PSF widths of 39 m at versus 11 m at for TR = 50 ms). For the flip-angle pair optimizing variable-flip-angle mapping, the effective resolutions differ by a factor of 2.4. In simulated DESPOT1 maps the of a 30 m feature is reduced to 1.80 s (true value 2.00 s). In b-SSFP, blurring depends on , , and but is independent of TR, with a minimum of . Conclusion: The PSF of a steady-state sequence depends on the flip angle and . Combining images acquired at different flip angles implicitly assumes identical spatial resolution, but that is not valid at microscopic scales, affecting the accuracy of quantitative relaxometry. The theory identifies a new limitation on the ultimate resolution of MRI set by diffusion.

生成讲解博客的中文版本

生成AI智能校验的中文版本