Reliable tool–tissue contact detection can support interaction-aware control and downstream force estimation in robot-assisted surgery. Most existing methods learn a contact classifier from RGB appearance, which is hard to generalize. In this work, we use the depth image generated from a stereo pair to give more information about tool–tissue contact. For each depth frame, we localize a spatially supported minimum-distance patch around the tool boundary and reduce it to a single scalar, ; this signal rises and falls in step with ground-truth contact. We formalize this observation with a fully supervised two-state hidden Markov model. We fit this model as a six-fold leave-one-session-out (LOSO) ensemble on six palpation sessions against a single silicone cup-like phantom, with the decision threshold selected from the pooled out-of-fold predictions. It is evaluated on four held-out sessions of three categories: 1. same task on same phantom; 2. same task on different phantom; 3. different task on different phantom. This model reaches held-out macro F1 and AUPRC . We further compare against a reproduction of an RGB-based contact classifier from prior work. This RGB-based model achieves high performance on the first category (F1 ), but substantially lower performance on the other two, resulting in macro F1 across all four sessions. These results indicate that the tool–tissue distance is a strong, transferable cue for contact detection in robot-assisted surgery.