Accurate Monte Carlo simulation of plastic scintillator detectors requires precise wavelength-dependent optical properties. We present the first comprehensive measurements of wavelength-dependent refractive index and bulk absorption length for six commercial polyvinyltoluene (PVT)-based scintillators (Saint Gobain BC-400, BC-404, BC-408, BC-418, BC-422, and BC-430) across the 370-660 nm range. Using a precision refraction apparatus and variable-thickness transmission measurements, we characterize material-specific dispersion that deviates substantially from both the manufacturer specification and literature values for pure PVT. Validation simulations of a BC-422 detector demonstrate a 7.2% reduction in detected photon counts when using our measured compared to the constant manufacturer value, highlighting the practical importance of accurate optical property data for detector design and calibration. We successfully measure absorption lengths for BC-430 across most of the visible spectrum (2.2-12.4 mm) and for all materials at 372 nm (21-38 mm), while establishing lower bounds exceeding 750 mm for weakly absorbing formulations at longer wavelengths.